"Generic" Andor cameras are detected before iXon 888 #2

@whooie

Description

@whooie

I wasn't able to work out why, but my camera was detected and loaded as "Generic Andor SDK2", and as a result I couldn't access any of the "Advanced" settings like trigger mode, shift speed, temperature control, etc.

I got around this by moving all of the definitions for AndorSDK2IXONCameraDescriptor (in utils/cameras/AndorSDK2.py) to the base class AndorSDK2CameraDescriptor, but other settings found in the Solis interface like "Acquisition Mode", "Readout Mode", and "Output Amplifier" (some others as well) are missing. Also, the options for "Horizontal shift speed", "Vertical shift period", and "Preamp gain" don't match the Solis interface.