Respirator Monitor System (RMS) is a software aimed to the RPM and pressure parameters of a respirator. It offers a web interface as well as a REST API in order to receive the information of the device.
The RMS will be connected to a magnetic switch on the respirator's arm in order to monitor its movement as well as report the RPMs. The RMS is equiped with an alarm to alert the user.
The following events will trigger the alarm (by order of priority): 1. If in a time difference of MAX_DIFF_SAMPLES no sensor new interruptions are triggered. This means that the motor has stopped. It is important to set up the MAX_DIFF_SAMPLES value taking into account the worst case scenario which corresponds to the cicle duration in MIN_RPM_MOTOR. 2. If the motor module reports an rpm value of -2. This code represents an internal error. 3. If the RPM value is not in the range of [MIN_RPM_MOTOR, MAX_RPM_MOTOR] 4. Any other error with the reading of the RPM value
- Motor monitor
- Pressure sensor monitor
- Grab the image file from here
- Flash the image to an SD card with Raspberry's utility
- Plug the SD card to the Raspberry and connect the alarm to pin 15, the stop alarm button to pin 13 and the motor sensor to pin 11. Both of this last two components must be low level active.
- The current version does not have a configured WiFi network. In order
to enable wireless connection, first of all access to the terminal via
sshwith a wired connection or directly with the terminal. Runsudo nano /etc/wpa_supplicant/wpa_supplicant.confand field thessidandpskfields. Do not forget to include the surrounding "". - Reboot the Raspberry and enjoy!
This is an example of the config, it is changed at __init__.py:
config = {
"Respirator": {
"ID": hex(getnode())[2:].upper()
"LOC": "Not set",
"POLL_FREQ": 1
},
"Motor": {
"STARTUP_TIME": 60,
"MIN_RPM_MOTOR": 12,
"MAX_RPM_MOTOR": 35,
"MAX_DIFF_SAMPLES": 6
}
}
- ID: Indicates the identification number of the respirator.
- LOC: Indicates the localization of the respirator.
- POLL_FREQ: Respirators control poll frequency.
- STARTUP_TIME: Minimum time to gather samples before monitoring.
- MIN_RPM_MOTOR: Minimum operating RPM.
- MAX_RPM_MOTOR: Maximum operating RPM.
- MAX_DIFF_SAMPLES: Maximum time difference between two consecutive samples. It is adviced to use the MIN_RPM_MOTOR cicle duration to take into account the worst case scenario.
In order to retrieve the information of the respirator the api link is the following:
http://example.com/api/status
Here there is an example of the output:
{
"id": "123",
"loc": "SF34",
"rpm": 40,
"status": "on"
}
- id: Identification number of the respirator.
- loc: Localization of the respirator.
- pressure: Pressure of the respirator.
- rpm: RPMs of the respirator.
- status: Status of the respirator. Can be
offwhen nothing has been received, ,calwhen it receives the first sample and suring the set up interval,onwhen functioning correctly orfailin case of faliure.
In order to change the location parameter, a POST request is needed. The body
of it must be:
{"loc": "new_loc"}
It will return the same information as with the GET /api/status if the
parameter was correctly change or {"error": "Location not changed"} if it
fails to do so.